    Title: Accuracy of point-of-care blood glucometers in neonates and critically Ill adults
    Authors: Sheen, YJ;Wang, JM;Tsai, PF;Lee, WJ;Hsu, YC;Wang, CY;Sheu, WH
    Contributors: Institute of Molecular and Genomic Medicine
    Abstract: PURPOSE: Inpatient glycemic management has become a common issue because of the increasing number of hospitalized patients with hyperglycemia. Point-of-care devices can enable timely inpatient glucose monitoring, which may lead to better outcomes. The accuracy of point-of-care testing in various clinical scenarios has been questioned, particularly in neonates and critically ill patients. This study aimed to evaluate the accuracy of the CONTOUR PLUS and CONTOUR PLUS ONE glucometers (new wireless systems that link to a smart mobile device) when used as point-of-care devices for blood glucose monitoring in neonates and critically ill adults in inpatient settings. METHODS: This cross-sectional study was conducted at a medical center in central Taiwan and enrolled patients admitted to the neonatal intensive care unit, sick child room, or respiratory intensive care unit between November 2020 and April 2021. Neonates with suspected infection or abnormal blood coagulation and adults who had abnormal blood coagulation, were pregnant, had received organ transplants, or had undergone massive blood transfusions were excluded. The accuracy of the glucometers was determined based on the following criteria of the International Organization for Standardization (ISO) standard: 15197:2013. FINDINGS: Overall, 114 neonates (mean age, 4.2 days [range, 0-28 days]; 65 boys [57.0%]) and 106 hospitalized critically ill adults (mean age, 68.2 years [range, 27-94 years]; 72 men [67.9%]) were enrolled in this study. The glucose values obtained with each glucometer had good precision, and all findings met the reference criteria of the within-lot results. All measurements of the neonates' venous blood by each glucometer met the accuracy criteria specified by ISO standard 15197:2013. Furthermore, 98.1% and 97.2% of the arterial blood glucose measurements for critically ill adults obtained with CONTOUR PLUS and CONTOUR PLUS ONE met the accuracy criteria, respectively. IMPLICATIONS: Both glucose management systems met the accuracy criteria for venous blood from neonates and arterial blood from critically ill adults. Thus, the use of these 2 point-of-care devices in inpatient settings, including for neonates and critically ill adults, can be recommended to minimize limitations associated with the clinical application of point-of-care testing in glucose management. The wireless connection may play a role in the subsequent development of institution-wide virtual glycemic management under the supervision of a team of endocrinologists.
    Date: 2023-07
    Relation: Clinical Therapeutics. 2023 Jul;45(7):643-648.
    Link to: http://dx.doi.org/10.1016/j.clinthera.2023.05.005
    JIF/Ranking 2023: http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=NHRI&SrcApp=NHRI_IR&KeyISSN=0149-2918&DestApp=IC2JCR
    Cited Times(WOS): https://www.webofscience.com/wos/woscc/full-record/WOS:001067014000001
    Cited Times(Scopus): https://www.scopus.com/inward/record.url?partnerID=HzOxMe3b&scp=85160331390
    Appears in Collections:[Huey-Herng Sheu] Periodical Articles

    Files in This Item:

    File Description SizeFormat
    PUB37248091.pdf687KbAdobe PDF105View/Open


    All items in NHRI are protected by copyright, with all rights reserved.

    Related Items in TAIR

    DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library IR team Copyright ©   - Feedback