國家衛生研究院 NHRI:Item 3990099045/13001
English  |  正體中文  |  简体中文  |  全文笔数/总笔数 : 12145/12927 (94%)
造访人次 : 853158      在线人数 : 668
RC Version 6.0 © Powered By DSPACE, MIT. Enhanced by NTU Library IR team.
搜寻范围 查询小技巧:
  • 您可在西文检索词汇前后加上"双引号",以获取较精准的检索结果
  • 若欲以作者姓名搜寻,建议至进阶搜寻限定作者字段,可获得较完整数据
  • 进阶搜寻
    主页登入上传说明关于NHRI管理 到手机版


    jsp.display-item.identifier=請使用永久網址來引用或連結此文件: http://ir.nhri.org.tw/handle/3990099045/13001


    题名: Use of the Viabahn covered stent for the treatment of venous rupture during interventions of dysfunctional or thrombosed hemodialysis vascular access
    作者: Lee, CK;Hsieh, MC;Luo, CM;Liao, MT;Hsieh, MY;Wu, CC
    贡献者: Institute of Cellular and Systems Medicine
    摘要: Background: Angioplasty-related vessel rupture is a common complication of interventions. The effect of covered stents to treat venous rupture has been evaluated in smaller series, but should be further evaluated. Objective: To report the immediate outcomes and patency rates of a covered stent to rescue angioplasty-related venous rupture of hemodialysis vascular access. Methods: From January 2013 to December 2018, 113 procedures complicated with vessel ruptures were retrospectively analyzed from a prospectively collected database of 8146 hemodialysis access interventions. The strategies to salvage vessel ruptures were based on the discretion of the treating physicians. Follow-up outcomes were obtained via review of the angiographic images, procedural notes, and medical and dialysis records within 12 months after the index procedures. Results: A total of 52 vessel ruptures (21 fistulas, 31 grafts) salvaged by using Viabahn covered stents were enrolled. Vessel ruptures developed in 28 (53.8%) thrombectomy procedures. Device success was achieved in all procedures (100%) and clinical success was achieved in 50 (96.2%). The primary patency of the stent area was 66.0% at 6 months and 50.0% at 12 months. The primary patency of the entire access circuit was 27.4% at 6 months and 16.0% at 12 months. The most common cause of access circuit primary patency loss was thrombotic occlusion for graft accesses and restenosis at stent area for native accesses. Eleven vascular accesses were abandoned within 12 months after vessel ruptures, and the secondary patency rate of the entire access circuit was 78.0% at 12 months. Conclusions: Treatment of angioplasty-induced vessel rupture of hemodialysis vascular accesses by using Viabahn covered stents has good immediate outcomes and patency results at the stent area. Nonetheless, the patency rate of entire access circuit was still below the threshold recommended by guidelines.
    日期: 2021-09
    關聯: Journal of Vascular Access. 2021 Sep;22(5):759-766.
    Link to: http://dx.doi.org/10.1177/1129729820961955
    JIF/Ranking 2023: http://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=NHRI&SrcApp=NHRI_IR&KeyISSN=1129-7298&DestApp=IC2JCR
    Cited Times(WOS): https://www.webofscience.com/wos/woscc/full-record/WOS:000575454900001
    Cited Times(Scopus): https://www.scopus.com/inward/record.url?partnerID=HzOxMe3b&scp=85092088691
    显示于类别:[吳志成] 期刊論文

    文件中的档案:

    档案 描述 大小格式浏览次数
    ISI000575454900001.pdf590KbAdobe PDF231检视/开启


    在NHRI中所有的数据项都受到原著作权保护.

    TAIR相关文章

    DSpace Software Copyright © 2002-2004  MIT &  Hewlett-Packard  /   Enhanced by   NTU Library IR team Copyright ©   - 回馈